Voltage Calibrators : Krohn Hite MV106

此产品是工厂翻新

Krohn-Hite MV106 DC Voltage Calibrator
放大图片Krohn-Hite MV106 DC Voltage Calibrator

Call 1-800-615-8378 for more information

Request a Quote

Check Rental Availability

省钱-购买二手设备


The Models MV106 is a precision voltage standards/sources that are highly versatile references designed to meet the needs of computer systems, production line testing, automated calibration systems and standards laboratories.  The MV106 provides 3 voltage ranges which include 10mV, 100mV and 10V.  Resolution is 0.01 V.

SPECIFICATIONS

Output: 10V, ±11.11110Vdc; 100mV, ±111.1110mVdc; 10mV, ±11.11110mVdc
Resolution (1ppm):
10V, 10 V; 100mV, 0.1 V; 10mV, 0.01 V
Output Accuracy (2 Methods) Calibration Accuracy:
±(0.003% of setting + 0.0005% of range + 2 V); at standard conditions at time of calibration; nominal temperature of 23 C ±1 C at <70% relative humidity; nominal line voltage 117Vac at constant external load.
Limit-of-Error (Worst Case Method):
±(0.005% of setting + 0.00075% of range + 2 V). This method includes all peripheral, additive specification, (i.e., error for line and load changes, ambient temperature changes and stability). Calibration cycle, 1 full year.

Stability (non-additive):
1hr., ±(0.0005% of range + 2 V); 8hrs., ±(0.001% of range + 2 V); 1yr., ±(0.0025% of range + 2 V)
Ripple & Noise, rms, 0.8Hz to 100kHz (non-additive):
10V, 50 V; 100mV, 1 V; 10mV, 0.5 V
Output Current (with remote sensing) See Output Impedance:
10V, 50mA; 100mV,
into RL>100k ohms; 10mV, into RL>100k ohms
Line & Load Regulation (non-additive):
±0.0005%
Output Impedance (All Zout are fixed constants):
10V, 0.03 ohms; 100mV, 3 ohms; 10mV, 3 ohms

GENERAL
Temperature:
Calibration, 23 C ±1 C; ambient, 20 C to 30 C; operating limit, –10 C to 50 C; storage, –40 C to 85 C.
Temperature Coefficient: Ambient, ±0.0005%/ C; operating limit, ±0.001%/ C.
Settling Time: Step change, 300ms; range change, 700ms.
Common Mode Rejection: 120dB, dc to 400Hz.
Isolation: Either terminal may be floated 500Vdc with respect to chassis ground; 1G ohm, 20pF.
Power Requirements: Selectable 115V or 220Vac ±10%; line frequency, 50Hz to 400Hz; power, 5 watts.
Circuit Condition Indicator: Front panel indicator for short-circuit, overload, over-voltage condition, low line voltage or malfunction.
Protection: Short-circuit and overload protection automatic recovery, over voltage protection up to 150Vdc.
Dimensions: 5.22" (13.3cm) high, 17" (43.2cm) wide, 10.9" (27.7cm) deep.
Weight: 13 lbs. (6.0kg) net; 17 lbs. (7.7kg) shipping.

MISCELLANEOUS
Certification:
Certificate of Compliance traceable to the U.S. N.I.S.T.
Warranty: One year, includes accuracy and stability.
Evaluation: Instruments are available for engineering evaluation. Contact our local sales representative or the factory for more information.

  • 制造商: Krohn Hite
  • 模型: MV106