Precision Magnetics Analyzer 20Hz-3MHz



  • Wide frequency range of 20 Hz to 3 MHz

  • Fast measurement speed - up to 20 measurements per second

  • 0.1% basic accuracy

  • Up to 125 A of DC bias current

  • Telecom measurement functions

  • Analysis mode with graphical displays

  • Comprehensive measurement functions including leakage tests

  • Straightforward intuitive operation

  • Print test results

  • IEEE 488GPIB control with LabVIEW driver



    At the design stage of component development it is vitally important to understand how the component performs under different operating conditions. This might include operation at diverse frequencies, AC drive levels or DC bias currents.



    The 3260B Precision Magnetics Analyzer can plot any of the measurement functions, such as inductance (L) or impedance (Z), (including secondary term) against frequency, AC drive level or DC bias current.



    Frequency sweeps within the range 20 Hz to 3 MHz can be selected. There is a choice of either linear or logarithmic frequency displays.



    3260B is especially suitable for measuring parameters on telecom transformers.



    The selected measurement parameter and its secondary value are presented graphically. AC drive levels can be set between 1 mV and 10 V. DC bias current can be set from 1 mA to 1 A internally. Using external 3265B 25 A DC Bias Units bias currents can be set to a maximum of 125 A.

    Measure Insertion Loss and Return Loss on telecom transformers



    With the dramatic growth of PCs connected to the telephone system for Internet access has come the requirement to measure Insertion Loss (IL) and Return Loss (RL) of line matching transformers.



    The 3260B Precision Magnetics Analyzer not only measures IL and RL has this capability. but the instrument also allows the user to enter the values for terminating resistance or impedance, if complex, and to select a damped network or blocking capacitor if required.

    Specification summary

    Measurement functions Z, , L, C, Rac, Rdc, Q, D

    Turns Ratio

    Primary and secondary Leakage Inductance

    Interwinding Capacitance

    Resonant Frequency

    Frequency ranges 20 Hz to 3 MHz

    Basic accuracy 0.1%

    Mode Analysis (graphing)

    Telecom

    Multi Frequency

    Sequence

    DC bias current 1 mA to 1 A (internal)

    Up to 125 A (using five 3265B DC Bias Units)

    Interface GPIB

    Measurement speed Up to 20 measurements/sec

    Printed output of test results



    Using the parallel Centronics interface the user can directly print test results including graphs for further analysis and archiving.



    In addition, via the IEEE488 GPIB interface, the instrument can be controlled from a PC and results can be read back for analysis and storage. LabVIEW drivers are available upon request or may be downloaded from this website providing a base from which a user can develop a specific test application.

    Bin sort



    The bin sorting function allows component manufacturers to sort components in up to ten bins. Sorting is carried out either by absolute values or by percentage of values.